Tablet Disintegration Tester - Qinsun
Professzionális vizsgálóműszerek gyártója
Nemzeti konzultációs hotline:+86 021-64208466
Termékek

Tablet Disintegration Tester

Meets current specifications of USP, IP, and other Pharmacopoeias

WhatsApp: +86 13816217984

Email: info@qinsun-lab.com

Tartalom

Jellemzők:

Meets current specifications of USP, IP, and other Pharmacopoeias

Specifically designed for use in the quality and production control of normal plain coated tablets and gelatin capsules

Programmable Temperature and Time

Molded, Clear acrylic water bath with illumination for bettervisibility

Baskets are automatically lifted at the end of the test

Capability for dual buffer disintegration test

Audible, low water level alarm with indication on display for safety.

External temperature sensors available for individual beaker


Description:

The Tablet Disintegration Tester is advance microcontroller based with user friendly software in dialog mode. The Alpha Numeric splash proof keyboard allows entries such as samplename, sample number or ID number for data authentication.The report printout complies with GLP requirements and abuilt-in Real Time Clock provides date, time display, and authenticity of the printout.


Műszaki adatok:

Operating Modes: Count or Time, user selectable

Dip Speed: 30 ± 1 DPM

Stroke Length: 5.5 ± 0.1 cm

Water Bath Temperature Range: ambient +5 to 50°C

Water Bath Temperature Accuracy: ± 0.2°

Water Bath Circulation: Immersion Pump

Beaker Volume: 1000 mL

Basket: 2 x 6 test positions (10 mesh) USP Type A

Fluted Disc: 6 in each basket (USP Type)

Test Mode:

Programmable Time: Up to 99:59 (hh:mm)

Display Format: 99:59:59

Optional: Dual Station with individual drive available

Display: 20 x 2 Lines, Back light LCD Display

Keyboard: Alphanumeric, splash water proof, polyester soft keys

Output: Printer, RS-232C, Parallel Port, PC Connectivity for Data Downloading

Environmental Operating Conditions:

Operation: Indoor

Temperature: Ambient to 45°C

Humidity: 20% to 80%

Electrical Requirements:115V/230V


Test Method:

This test is provided to determine whether tablets or capsules disintegrate within the prescribed time when placed in a liquid medium at certain experimental conditions.



Felelősségi nyilatkozat: A fenti tartalom kizárólag az iparági belső személyek közötti hivatkozásra és kommunikációra szolgál, és nem garantálja pontosságát vagy teljességét. A vonatkozó törvények és előírások, valamint a weboldal előírásai szerint a kapcsolódó cikkeket vásárló egységeknek vagy egyéneknek érvényes képesítést és képesítési feltételeket kell megszerezniük.