STD-TE 60 HIGH PRESSURE HYDROGEN RECIPROCATING TRIBOMETER - Qinsun
Professzionális vizsgálóműszerek gyártója
Nemzeti konzultációs hotline:+86 021-64208466
Termékek

STD-TE 60 HIGH PRESSURE HYDROGEN RECIPROCATING TRIBOMETER

DescriptionSTD-TE 60 is a three-station reciprocating tribometer, designed around a high pressure c

WhatsApp : +86 13816217984

Email : info@qinsun-lab.com

Tartalom

Leírás

STD-TE 60 is a three-station reciprocating tribometer, designed around a high pressure chamber, for experiments under hydrogen and other gases. It includes a vacuum pump for purging the chamber before charging with a chosen gas, so can also be used for tests under medium vacuum. Tests may be run with plate specimen temperatures from -50°C to +150°C.

Technical Specifications

Technical Specifications


Number of Test Stations:Three
Test Configuration:Ball on Flat

Cylinder on Flat
Gases:Hydrogen, Argon, CO2
Working Pressure:0.1 mbar to 100 bar
Proof Pressure:120 bar
Stroke Range:0 to 20 mm Continuously Adjustable
Frequency:0 to 5 Hz
Motion:Sinusoidal
Loading System:Tension Spring
Load Range:5 to 100 N
Friction Range:0 to 50 N
Deflection Sensor:Inductive Transducer
Pressure Rating:150 bar
Refrigerant Probe:
Action:Contact cooling of plate specimen
Temperature Range:Ambient to -50°C
Heating Probe:
Action:Contact heating of plate specimen
Temperature Range:Ambient to 150°C
Electrical Feedthroughs:PT100 x 2
Rotary Feedthough:BMD-300 Magnetic Coupling
Motor:NX-210 a.c. Servo Motor


Automatically Controlled Parameters
Frequency
Plate specimen temperature
Test duration


Manually Controlled Parameters
Load
Stroke
Chamber pressure (external service)


Measured Parameters
Friction
Plate specimen temperature
Chamber temperature
Chamber pressure
Number of cycles


Derived Parameters
Friction coefficient


Services
Electricity:220/240V, single phase 50/60 Hz, 7.5 kW
Process Gas System:Not included


ATex RatingThe machine, drive motor and associated instrumentation are ATex rated. The chiller unit, vacuum pump, power module and PC control and data acquisition system are not ATex rated. It follows that these items must be installed outside any ATex rated test cell, with suitable ATex rated connections between the operator work-station and the test cell.


Felelősségi nyilatkozat: A fenti tartalom kizárólag az iparági belső személyek közötti hivatkozásra és kommunikációra szolgál, és nem garantálja pontosságát vagy teljességét. A vonatkozó törvények és előírások, valamint a weboldal előírásai szerint a kapcsolódó cikkeket vásárló egységeknek vagy egyéneknek érvényes képesítést és képesítési feltételeket kell megszerezniük.