Total Hip Joint Prostheses Wear Tester - Qinsun
Professzionális vizsgálóműszerek gyártója
Nemzeti konzultációs hotline:+86 021-64208466
Termékek

Total Hip Joint Prostheses Wear Tester

- Evaluating the wear - resistance of hip joint prostheses: By simulating human movement states such

WhatsApp : +86 13816217984

Email : info@qinsun-lab.com

Tartalom

Standard

YY/T 0651.1-2016 (China): Specifies wear testing parameters for total hip prostheses, including angular motion, loading modes, and environmental conditions.

ISO 14242-1:2014 (International): Defines protocols for simulating physiological loading and kinematics during hip joint wear testing.

ASTM F3047M-15: Additional compliance for advanced testing scenarios including edge loading and micro-separation.

Product Advantages

Triaxial Motion Simulation: Replicates full physiological motion profiles (flexion/extension ±100°, abduction/adduction ±30°, internal/external rotation ±40°).

Dynamic Load Control: Dual-peak loading up to 4.500 N with 0.1% force accuracy.

Biorelevant Environment: Temperature-controlled (37±2°C) lubrication bath with 30±2 g/L protein concentration.

Extended Durability: Capable of 5+ million cycle testing with <0.5% drift in calibration.

Automated Compliance: Pre-programmed ISO/YYT test protocols with real-time data validation.

Modular Design: Six-station configuration allows parallel testing of multiple specimens.

Product Features

Electrohydraulic Actuation: Six independent servo-hydraulic actuators for precise motion control.

3D Force/Torque Sensors: Multi-axis sensors measure axial forces and torsional moments simultaneously.

Temperature-Controlled Bath: Recirculating lubrication system with protein concentration monitoring.

High-Resolution Imaging: Optional integration with micro-CT systems for in-situ wear analysis.

AI-Powered Analytics: Machine learning algorithms predict wear rates from initial test cycles.

Safety System: Emergency stop, over-pressure relief, and laser-based specimen breakage detection.

Műszaki paraméterek

ParaméterSpecifikáció
Angular Motion RangeFlexion/Extension: ±100°
Abduction/Adduction: ±30°
Rotation: ±40°
Max. Axial Force±4.500 N (1.000 N resolution)
Load Control Accuracy±0.5% of reading (0–4.500 N)
Frequency Range0.1–5 Hz (programmable waveforms)
Temperature Control37±2°C (PT1000 sensors)
Lubrication System15L capacity with protein skimmer and bubble trap
Data Acquisition100 kHz sampling rate (16-bit resolution)
Power Requirements220V/110V ±10%, 50/60Hz, 2.500W
Dimensions (W×D×H)1.800 × 1.200 × 2.000 mm
Súly950 kg (including bath and control cabinet)
Noise Level<65 dB (A-weighted)

Standard Accessories

Hip Simulator Fixture Kit:

Self-centering femoral head adapters (16–32 mm)

Acetabular cup positioning jigs (60° inclination standard)

Biorelevant Bath System:

15L capacity with recirculation pump and protein skimmer

316L stainless steel construction with HEPA filtration

WearMaster Pro Software:

Automated wear rate calculation (mass loss/volume displacement)

21 CFR Part 11-compliant audit trail with blockchain timestamping

Remote monitoring via iOS/Android app

Operator Kit:

Calibrated torque wrench set (1–50 Nm)

Specimen alignment gauges (ISO 14242-1 compliant)

Calibration Bundle:

NIST-traceable force calibration weights

Angular encoder verification tool

Safety Kit:

Laser safety curtains

Acoustic enclosure with 30 dB noise reduction

Felelősségi nyilatkozat: A fenti tartalom kizárólag az iparági belső személyek közötti hivatkozásra és kommunikációra szolgál, és nem garantálja pontosságát vagy teljességét. A vonatkozó törvények és előírások, valamint a weboldal előírásai szerint a kapcsolódó cikkeket vásárló egységeknek vagy egyéneknek érvényes képesítést és képesítési feltételeket kell megszerezniük.