I. Product Overview The Horizontal Combustion Tester is a standardized instrument designed to evalu

WhatsApp: +86 13816217984

Email: info@qinsun-lab.com

Tartalom

I. Product Overview

The Horizontal Combustion Tester is a standardized instrument designed to evaluate the flame propagation characteristics of textiles, plastics, and composite materials under controlled horizontal flame exposure. It quantifies ignition resistance, flame spread rate, and self-extinguishing behavior in compliance with international fire safety standards.

  • Type: Horizontal flame propagation tester
  • Test Standards: ASTM D6413, ISO 6941, NFPA 701
  • Key Metrics:
    • Flame Spread Time (seconds)
    • Char Length (mm)
    • Afterflame Duration (seconds)
  • Sample Size: 610×1220mm (horizontal clamping)
II. Technical Specifications
Paraméter Specifications
Burner Type Propane gas burner (12mm ceramic nozzle)
Flame Height 125±5mm (adjustable via gas flow)
Test Duration 10–300 seconds (programmable)
Flame Propagation Sensor High-speed thermal camera (120Hz)
Tápegység 220V/50Hz, 2.5kW
Méretek 1500×800×1800mm (L×W×H)
III. Key Features
  • Dual Test Modes:
    • Horizontal Test: Standard flame spread evaluation
    • Edge Stress Test: Simulates fabric edge ignition
  • Advanced Detection:
    • Real-time thermal imaging (°C gradient mapping)
    • AI-based flame front tracking
    • Automatic char length measurement (±1mm accuracy)
  • Automation:
    • Motorized burner positioning
    • Programmable gas flow control
    • Self-cleaning exhaust system
IV. Compliance Standards
  • ASTM D6413 (Vertical and horizontal flame propagation)
  • ISO 6941 (Textile flame spread rate)
  • NFPA 701 (Standard for flammability of draperies)
  • GB/T 5455 (Chinese textile combustion standard)
V. Applications
  1. Textile Manufacturing: Certifying fabrics for public buildings
  2. Interior Design: Evaluating drapery fire resistance
  3. Fire Safety Departments: Compliance testing for hotels/hospitals
  4. R&D: Developing flame-retardant coatings
Standard Operation Workflow:
  1. Sample Preparation: Cut to size, clamp horizontally
  2. Calibration: Reference fabric validation
  3. Testing: Ignite burner at specified height/duration
  4. Analysis: Generate flame spread velocity curves
  5. Cleaning: Automated soot extraction system
Safety Precautions:
  • Fire-rated enclosure required (class A60)
  • PPE mandatory (heat-resistant gloves, goggles)
  • Fire extinguisher (Class B) must be on standby
* ISO 17025-accredited testing protocol * Multi-layer flame front analysis * Real-time audio-visual alarm system * Gas leak detection with auto-shutdown * Includes 5-year burner maintenance ​**Technical Highlights**​ - ±0.5s timing accuracy (NIST-traceable chronometer) - 360° thermal imaging coverage - Automatic gas pressure compensation - Touchscreen data export via USB/4G - OSHA-compliant emergency stop protocol

image.png

Felelősségi nyilatkozat: A fenti tartalom kizárólag az iparági belső személyek közötti hivatkozásra és kommunikációra szolgál, és nem garantálja pontosságát vagy teljességét. A vonatkozó törvények és előírások, valamint a weboldal előírásai szerint a kapcsolódó cikkeket vásárló egységeknek vagy egyéneknek érvényes képesítést és képesítési feltételeket kell megszerezniük.


Online üzenet

Kapcsolat adatai

Vállalati telefon

+86-21-6420 0566

munkaidő

Hétfőtől péntekig

Mobiltelefon:

13816217984

E-mail:

info@qinsun-lab.com

QR kód