Measures gas released by materials upon contact with water.
WhatsApp: +86 13816217984
Email: info@qinsun-lab.com
Product Description
GB/T 21619-2008 Test Methods for Dangerous Goods Flammable Solids Emitting Flammable Gases in Contact with Water, GB 19521.4-2004 Safety Standard for Hazardous Characteristics Inspection of Dangerous Goods emitting flammable gases in contact with water, GA/T 536.4-2005 Fire Hazard Classification and Test Methods for Flammable and Explosive Hazardous Goods, Part 4: Substances emitting flammable gases in contact with water, GB5085.5-2007 Hazardous Wastes Identification Standard for Reactive Identification, and NY/T 1860.5-2007 Hazardous Waste Identification Standard for Reactive Identification. and NY/T 1860.29-2016 Test Guidelines for the Determination of Physical and Chemical Properties of Pesticides Part 29: Flammability in Contact with Water and other standards for testing.
Product features
1) Adopting pressure balance method, with high-precision pressure detection and precision motion control, it can effectively collect the gas released by the substance in contact with water, and accurately calculate the release rate and the amount of gas.
2)Simultaneously display the maximum release rate in minutes and hours and the gas collection volume, and draw the rate curve.
3)Traceable instrument calibration function and calibration tools to ensure the accuracy of the instrument.
4)The sample is automatically filled with water under closed conditions to avoid gas leakage during water filling and sample addition.
5)The thermostatic circulation tank and the special metal plate realize the heat and cold conduction to maintain the constant temperature of the test chamber.
6) The gas pipeline is well sealed and has the ability to resist acid and alkali corrosion.
7)With automatic cleaning function of gas line to realize the rapid discharge of residual gas inside the instrument.
8) Configuration of gas collection bag to automatically collect the released gas, in order to analyze the composition of the gas or combustion test
9)The door is equipped with observation window and illumination, which is convenient to observe the reaction state.
10)The accessories can be configured to carry out the water test, stay test and drip test.
11) The test process is fully automated, without human intervention, support for parameter setting, data storage, view reports and other functions, real-time display of the test status, the background automatically save the original data
12) 7-inch LCD touch screen, Chinese intelligent input method, simple and convenient operation
13)Adopting embedded processor and Windows CE operating system, fast, stable and reliable operation.
Technical parameters
Operating Environment | (5~40)℃, <85%RH |
Temperature Control Range of Test Chamber | (10~35)℃, default temperature control target: 20℃, can be set by software |
Temperature Control Stability | ±0.5℃ |
Test Water Volume | 400mL, the sample automatically into the water |
Sample Cup Volume | 20mL |
Gas Collection Volume | 0~800mL |
Gas Detection Resolution | 0.01mL |
Gas Collection Accuracy | ±0.5mL |
Sample Gas Release Rate Range | 0~100L/min/kg |
Gas Volume Sampling Interval | 1s |
Gas Release Rate Calculation Interval | 1min~1h, can be set by software Timing range 0~240h |
Timing Range | 0~240h |
Felelősségi nyilatkozat: A fenti tartalom kizárólag az iparági belső személyek közötti hivatkozásra és kommunikációra szolgál, és nem garantálja pontosságát vagy teljességét. A vonatkozó törvények és előírások, valamint a weboldal előírásai szerint a kapcsolódó cikkeket vásárló egységeknek vagy egyéneknek érvényes képesítést és képesítési feltételeket kell megszerezniük.